Chapter 41Questions and problems
This part gathers what the screens say when something is wrong, from the chapters of this manual as they now stand, so that a message can be found in one place. Each row names the chapter it comes from; the full detail is there.
41.1 Messages by area
Rows that two chapters share appear once, with both chapters named.
41.1.1 Signing in and the first screens
| The screen says | Meaning | What to do | Chapter |
|---|---|---|---|
| The sign-in did not succeed. | The server refused the sign-in. | Check the user name and password, and try again. | Getting started |
| The sign-in was not completed. | The second step could not be completed. | Press Back and sign in again. | Getting started |
| The request on the phone ended. Type the code, or sign in again. | The phone's request timed out. | Type the code, or sign in again. | Getting started |
| The two passwords do not match. | The two passwords differ. | Type them again. | Getting started |
| The password could not be set. | The recovery code or the user name was not accepted. | Check them, or ask the administrator. | Getting started |
| This language could not be loaded from the server, so the screen keeps its language. Check that the server is running, then try again. | The language could not be loaded from the server. | Check the server is running and try again. | Getting started |
| Nothing found. Press Enter to look for a patient by this on the order screen. | The search found nothing. | Press Enter to look for a patient by that text on the order screen. | Getting started |
| The save failed, so the step stays as it was. | The step could not be saved. | It stays as it was; try again. | Getting started |
41.1.2 Home and the management figures
| The screen says | Meaning | What to do | Chapter |
|---|---|---|---|
| The list could not be loaded. | The list could not be fetched from the server | Reload the page; if it persists check the connection | Home |
| The overview could not be read. It will be asked again in a few minutes. | The overview could not be read; it asks again after a few minutes | Wait, or reload | Home |
| The money of these branches is left out, because no exchange rate is written for its currency in that month: …. Write it in «Rates of exchange». | A branch's money is left out because no exchange rate is written for its currency in that month | Write the rate in Rates of exchange | Home |
| The services cannot be read right now. | The services could not be read | Reload later | Home |
| This card cannot be read right now. | One card cannot be read | Try again later | Home |
| Nothing waits for your decision now. | Nothing needs your decision | Nothing | Home |
41.1.3 Orders and patients
| The screen says | Meaning | What to do | Chapter |
|---|---|---|---|
| Cannot save: … | Something is missing | Read the reason given with it | New order |
| The order was not saved | The server refused the order | Read the reason shown, correct and save again | New order |
| The price could not be calculated: … | The price could not be worked out | Press Try again | New order |
| No test has the code or name «…»; nothing was added. | No test has that code or name | Check the spelling | New order |
| The test … (…) is in the catalogue but is not switched on in this laboratory; nothing was added. | The test is off in this laboratory | Open the catalogue if you may, or ask whoever manages it | New order |
| This phone number is one digit repeated: write the patient's real number, or choose «No phone» | The phone is one digit repeated | Write the real number or choose No phone | New order |
| This is not a phone number in the laboratory's country; a foreign number is written with + and its country code | The number is not valid in the laboratory's country | Fix it; a foreign number needs + and its country code | New order |
| The member or card number is required for this account type. | The account type needs a member number | Write it in the coverage details | New order |
| Nothing in the order changed to be saved. | The edit changed nothing | Nothing to save | New order |
| Some sample statuses were not applied. | Some samples were not updated | Check the samples screen | New order |
| The patient directory could not be loaded. | A connection or server fault | Reload; check the connection | Patients |
| No file matches «…». | Nothing has this text | Check spelling; choose All in the files' state | Patients |
| This patient may be registered already. | A file with the same details exists | Open that file, or save anyway if it is another person | Patients |
| The file could not be saved. | The server refused the save | Read the field notes and correct | Patients |
| The file is saved; the consent to marketing messages could not be. | The file saved but the consent did not | Open the file again and record the consent | Patients |
| The merge failed. Nothing was changed. | The server refused the merge | Check the two files and try again | Patients |
| The message could not be sent. | A fault in sending | Try again | Patients |
| The documents could not be read. | A connection fault | Reopen the tab | Patients |
| The file is larger than 10 MB. | The document is too large | Attach a smaller copy | Patients |
| Reading the orders… | The list is loading | Wait | Orders |
| The orders could not be read. | A fault reading the list | Use the retry control and check the connection | Orders |
| No order matches «…». | The search found nothing | Check the spelling, or clear the search | Orders |
| No order yet. / No order today. / No order in progress. | Nothing in that filter | Choose another filter | Orders |
| Choose an order from the list. | None is chosen | Click an order | Orders |
| No price — needs pricing | A test has no price | See 10.7.4 Task 4 — Price an order that waits for a price | Orders |
41.1.4 Samples
| The screen says | Meaning | What to do | Chapter |
|---|---|---|---|
| No sample with the code … | The code was mistyped, or no such sample exists | Scan again, or search by the order's number | Samples |
| The code could not be read. | The scan failed | Scan again | Samples |
| That could not be saved. | The server refused the step | Read the reason shown, correct and repeat | Samples |
| The samples could not be read. | A connection fault | Check the connection and retry | Samples |
| … stored samples are past their retention | Samples kept longer than the retention time | Dispose of them (6.7.1 Task 1 — Follow a sample from receipt to storage) | Samples |
| The move could not be made. | The board refused the move | Look at the tubes and try again | Samples |
| Write the time as 2026-10-03 10:52 | The collection time is not in the expected form (shown under the field) | Write it as in the example | Samples |
| Rejection is final; the server opens a re-draw by itself. | A notice before rejecting | Be sure, then reject | Samples |
41.1.5 Results and approval
| The screen says | Meaning | What to do | Chapter |
|---|---|---|---|
| No sample or order has the number … | The code was mistyped, or belongs to another branch | Scan again; check the branch filter | Results and approval |
| Not everything typed was saved | The server refused some values | Read the refusal beside each; correct and save | Results and approval |
| The approval failed; nothing was approved | A connection fault | Check the connection and try again | Results and approval |
| The server refused to approve some tests | A check failed on the server | See the red marks in the foot | Results and approval |
| The value … is above the most the analyzer measures | Beyond the measuring range | Confirm, or dilute and rerun | Results and approval |
| The previous one was zero, so no percent can be taken | Delta by percent impossible | Review the value by eye | Results and approval |
| needs permission to read patient files | Your role cannot see Previous | Ask the administrator if you need it | Results and approval |
41.1.6 Quality control, microbiology, clinical AI and accreditation
| The screen says | Meaning | What to do | Chapter |
|---|---|---|---|
| Quality control could not be loaded. | A fault reading the screen's data | Retry; check the connection | Daily QC |
| No active control lots. Add one from Lots. | No active lot | Add a lot (12.7.4 Task 4 — Manage the control lots) | Daily QC |
| The action failed. Try again. | The server refused a step | Read the reason in red and try again | Daily QC |
| Open violation: … results … are held in Results until it is resolved with a written action. | A failed run holds that device | Resolve it (12.7.3 Task 3 — Resolve a violation) | Daily QC |
| This lot expires … Prepare the next one. | Expiry within 45 days | Add the next lot (12.7.4 Task 4 — Manage the control lots) | Daily QC |
| Unreadable range: … | A stored measuring range cannot be read | Open the rule and write the range again (12.7.5 Task 5 — Set the automatic release rules per test) | Daily QC |
| The replay failed. | The replay could not run | Narrow the period and try again | Daily QC |
| The replay reached its limit of results | Too many results in the period | Narrow the period | Daily QC |
| The rules could not be read. | A fault reading the rules | Close and reopen the window | Daily QC |
| The controls could not be read. | A fault reading the mappings | Close and reopen the window | Daily QC |
| It could not be saved. | The server refused the save | Check the fields and try again | Daily QC |
| The cultures could not be loaded. | The list did not arrive. | Use the retry offered under the list. | Microbiology |
| No cultures waiting to start. | Every ordered culture has been started. | Nothing to do. | Microbiology |
| The action failed. Try again. | The server refused or did not answer; when the server gives its own reason, that is shown instead. | Press the button again; if it persists, tell the administrator. | Microbiology |
| Susceptibility is shown from the laboratory's newest results only, until the server returns the whole culture. | The culture's full record has not arrived yet. | Wait a moment; the full table replaces it. | Microbiology |
| The assistant read no susceptibility results from this file. | The file had nothing the assistant could read. | Try the text, or add the results by hand (13.6.2 Task 2 — Add susceptibility results). | Microbiology |
| No cultures here. | The list is empty, so there is nothing to show in the details. | Change the tag or the search. | Microbiology |
| Choose a culture to see its growth and susceptibility. | Rows exist but none is chosen (the screen normally chooses the first one). | Press a row. | Microbiology |
| The service is not switched on yet | The service is off in this laboratory. | A manager can switch it on in Settings. | Clinical AI |
| The drafts could not be loaded. | The drafts could not be loaded. | Press the retry button on the list. | Clinical AI |
| The action failed. Try again. | An approve, reject or remove did not go through. | Try again. | Clinical AI |
| Choose a draft to review. | No draft is selected. | Select a row. | Clinical AI |
| No drafts here. | The filter has no drafts. | Choose another chip. | Clinical AI |
| No movement of the balance yet. | The balance has not moved. | Nothing to do. | Clinical AI |
| The standard could not be read. | The standard could not be read. | Use the retry button; if it persists, check the connection to the server. | Accreditation and quality |
| The change could not be saved. | The change of standards was not saved. | Try again; the list stays as it was. | Accreditation and quality |
| The change could not be saved. | The function switch was not saved. | Try again, or ask the person who holds the right. | Accreditation and quality |
| The workbook could not be exported. | The workbook did not download. | Press the button again. | Accreditation and quality |
| This register is switched off in this laboratory: its entries stay readable, and none is added or changed. | The register is off. | Switch it on from the Functions tab if you may. | Accreditation and quality |
| The library cannot be reached now. | The library cannot be reached. | Import the standard from a file. | Accreditation and quality |
| The history could not be read. | The history of an item did not load. | Close and open it again. | Accreditation and quality |
41.1.7 Delivery
| The screen says | Meaning | What to do | Chapter |
|---|---|---|---|
| The orders ready to hand over could not be read. | The list could not be read | Press retry; check the connection | Delivery |
| This order's release is held (…): it is not published or sent until the person who approves results lifts it. | A release hold is on the order | Wait until the hold is lifted | Delivery |
| The laboratory's setting lets only a billing manager hand over a report with money left | The setting reserves this to billing managers | Collect the remainder, or ask a billing manager | Delivery |
| Write the receiver's name, and where the doctor works when he receives it. | No receiver is written | Write the name | Delivery |
| The hand-over could not be recorded. | The hand-over was not recorded | Try again | Delivery |
| This server does not record hand-overs yet; nothing was recorded. | This server cannot record hand-overs | Nothing was recorded; contact support | Delivery |
| Nothing was sent: the patient's file has no phone number and no e-mail. | The file has no phone and no e-mail | Complete the patient's file | Delivery |
| This patient's phone number needs confirming: it may now belong to someone else. Open the patient's file, confirm or correct the number, and send again. | The number may belong to someone else now | Confirm it in the patient's file, then send | Delivery |
| The patient has no WhatsApp number on file. | No WhatsApp number | Add it to the patient's file | Delivery |
| No link a patient can open exists yet. Send «WhatsApp with the file» so the report arrives as a file, or publish the laboratory's website. | No link exists yet | Send the file instead, or publish the link first | Delivery |
| The report's PDF is saved and the chat is open: drag the saved file into the chat, then press Send in WhatsApp. | The chat is open but the file is not attached | Drag the saved file in and press Send | Delivery |
| The report could not be published. | The link was not published | Try again | Delivery |
| The result link could not be stopped. | The link was not stopped | Try again | Delivery |
| The version could not be signed. | A version was not signed | Try again | Delivery |
| That could not be done. | The action on messages failed | Try again | Delivery |
41.1.8 Finance: treasury, prices and claims
| The screen says | Meaning | What to do | Chapter |
|---|---|---|---|
| The books are not started yet | No finance page can read anything | Ask whoever holds the finance settings to start the books | Treasury |
| It cannot close now | A shift, a recurring expense or a document blocks the month | Clear the reason listed, then close again | Treasury |
| Cash from the drawer needs an open shift | The expense was taken from the drawer with no shift open | Open a shift, or choose another method | Treasury |
| This is your request, so another manager decides it | You cannot decide what you asked | Ask another manager | Treasury |
| The consolidation is shown once the missing rates are written | A branch's currency has no rate for the month | Write the rate under Rates of exchange | Treasury |
| It could not be saved. Try again. | The server refused or the connection failed | Check the fields and the connection and try again | Treasury |
| The prices could not be loaded. | The prices could not be read | Check the connection and press retry | Prices and contracts |
| This field is required. | A required field is empty | Fill it | Prices and contracts |
| Coverage and patient share together must make 100%. | Coverage and patient share given together do not make 100% | Correct one, or give one alone | Prices and contracts |
| The end date is before the start date. | The dates are reversed | Correct them | Prices and contracts |
| Enter the amount above which pre-authorisation is needed. | Pre-authorisation above an amount without the amount | Type the amount | Prices and contracts |
| An offer's percent cannot exceed 100%. | An offer in percent is above 100 | Lower it | Prices and contracts |
| Could not save. | The server refused the save | Read the reason that follows and correct | Prices and contracts |
| The library cannot be reached now. | The library cannot be reached | Import from a file instead | Prices and contracts |
| The library is closed to this copy now. | The library is closed to this copy | Import from a file instead | Prices and contracts |
| The claims could not be loaded. | The list did not arrive. | Use the retry button, then check the connection. | Claims |
| The claim cannot be staged until what it lacks is completed. | The order is missing something from the list. | Complete each item, then save again. | Claims |
| This claim was already staged; it was not created twice. | The order already has this claim. | Find it in the queue. | Claims |
| The payer's channel did not take the claim; the reason is kept on it. | The send failed. | Read Last error, correct, send again or export. | Claims |
| This write-off or adjustment is your decision, so someone else approves it. | You recorded it. | Ask another person to review it. | Claims |
| Enter the amount transferred or the adjustment. | The transfer window has neither. | Type one. | Claims |
| Write the reason for the hold or deferral. | The hold has no reason. | Write it. | Claims |
| The file cannot be marked sent while evidence is missing. | The appeal lacks papers. | Tick the papers in hand, or keep the file as a draft. | Claims |
| The address starts with https://… | A channel address does not. | Type the full address as the payer gave it. | Claims |
| Resolve the blocking checks first | The stage is blocked. | Clear its blocking checks. | Claims |
41.1.9 Send-outs
| The screen says | Meaning | What to do | Chapter |
|---|---|---|---|
| WhatsApp did not open: no usable courier number, or the browser blocked the window. | The call was recorded but WhatsApp did not open: no usable courier number, or the browser blocked the window. | Allow pop-ups or use Copy the text. | Send-outs |
| The message is written; send it another way. “Courier called” is recorded only once WhatsApp is opened. | The laboratory has no courier number, so there is no WhatsApp button and the call is not recorded. | Send the text another way. | Send-outs |
| This laboratory is switched off: switch it on, then assign tests to it. | The laboratory is stopped. | Activate it first. | Send-outs |
| No active laboratory yet. Add one in the «External laboratories» tab. | No active laboratory. | Add one in the laboratories page. | Send-outs |
| The laboratory can be changed while the sample is in the building; once it has left, it cannot. | Too late to change. | Contact the laboratory. | Send-outs |
| Choose at least a test, a payer or a branch. | A rule needs a scope. | Choose a test, payer or branch. | Send-outs |
41.1.10 Attendance, follow-up, inventory, quality, marketing and the Market
| The screen says | Meaning | What to do | Chapter |
|---|---|---|---|
| The device could not be reached. | The device did not answer. | Check its address, port and that it is on; test again. | Attendance |
| The device's clock is … minutes away from the server's: set it on the device. | The device's clock differs from the server's. | Set the clock on the device. | Attendance |
| This device has not called yet. | A pushing device has not called yet. | Set the server's address in the device. | Attendance |
| Set the working days first | The employee has no schedule. | Give a schedule on Employees and schedules. | Attendance |
| Could not save. | The change was not saved. | Try again; read the message shown with it. | Attendance |
| No expense recorded in the period. | No expense is recorded in the chosen period. | Nothing is wrong; choose another period if you expected expenses. | Follow-up |
| Store purchases are not counted in one branch's report: a purchase order carries no branch. They are counted for the whole laboratory. | A single branch's report leaves out store purchases, because a purchase order carries no branch. | Choose The whole laboratory to see them counted. | Follow-up |
| The file could not be downloaded. Please try again. | The Excel file did not download. | Press Export to Excel again; if it persists, check the connection to the server. | Follow-up |
| Not enough stock: … only. | A movement takes more than is in the store. | Lower the quantity or check the store. | Inventory |
| A transfer needs two different stores. | From and to are the same. | Choose two stores. | Inventory |
| Choose the category or store to count. | A count by category or store needs its scope. | Fill the category or store. | Inventory |
| Enter the lot of «…». | A received line has no lot. | Write the lot. | Inventory |
| «…» exceeds what was ordered beyond the over-receipt limit. | The quantity received is too high. | Lower it to the allowed level. | Inventory |
| Approved by someone with procurement approval who neither created nor sent it. | You may not approve your own order. | Ask another approver. | Inventory |
| Write the reason. | Cancel, close, refuse, quarantine and recall need a reason. | Type it. | Inventory |
| The tests linked to this item could not be read… | The item's test list failed to load. | Reopen the window. | Inventory |
| The import failed. Try again. | The file could not be imported. | Check the file and repeat. | Inventory |
| The records could not be read. | The records could not be read. | Press retry; if it persists, check the connection and your permissions. | Quality |
| Write the resolution before closing. | The close window was saved empty. | Write the resolution. | Quality |
| Without the assigned value and the standard deviation the result is not evaluated. | The assigned value or the deviation is missing. | Enter both if you want the round evaluated. | Quality |
| Choose a record from the list to see its details. | No record is chosen. | Choose one in the list. | Quality |
| Reading… / Could not be read. | The list is loading, or the server refused. | Wait, or press «Try again» beside the message, and check your connection. | Marketing |
| Nothing matches «…». | The search found nothing. | Change the words, or choose «All». | Marketing |
| No approved template for this channel | The channel has no approved template. | Approve one in the Messages page of the settings. | Marketing |
| Name the send / Choose the template / Enter the test code | The button is dim for this reason. | Fill the named field. | Marketing |
| A package has at least two tests. | A package was set with fewer. | Add tests. | Marketing |
| A package cannot cost more than its tests in the catalogue | The price is above the total of the tests. | Lower the price. | Marketing |
| The send failed. | The server refused the send. | Read the reason shown above the button and try again. | Marketing |
| 3 messages handed over for sending | The send was accepted. | Nothing; read the count of those without a number or who withdrew. | Marketing |
| This release does not carry the Market's content key yet, so no document is accepted until an update brings it. | This release cannot yet accept documents. | Wait for an update that brings the key. | Clarus Market |
| Clarus has published nothing in the Market yet: there is nothing to install for now, and your connection is working. | Nothing to install now. | Nothing; try later. | Clarus Market |
| The request could not be completed. | The request did not complete. | Check the connection and try again. | Clarus Market |
| Nothing of this list was applied. | No row of this list was applied. | Nothing. | Clarus Market |
41.1.11 Devices and analyzers
| The screen says | Meaning | What to do | Chapter |
|---|---|---|---|
| No analyzer is registered in this laboratory yet. | No analyzer has been added. | Press Add an analyzer. | Devices and printers |
| No link open now | The analyzer is not connected at this moment. | Check the cable or network, then Test the connection. | Devices and printers |
| The analyzer has sent no messages yet. | Nothing has arrived from the analyzer yet. | Send a test from the analyzer, or check the settings it was given. | Devices and printers |
| This code's results reach no order until it is mapped to a test of the catalogue. | The code is not mapped. | Map it (31.9.3 Task 3 — Map a code to a test). | Devices and printers |
| No station serves this analyzer to test the connection: run the program on the computer it is connected to. | No station serves this analyzer. | Run the program on the computer it is connected to. | Devices and printers |
| No station has been marked as cabled to this analyzer yet: tick it on its station in «Stations and printers». | No station is marked as cabled to it. | Tick it on its station (31.9.8 Task 8 — Cable an analyzer to a station). | Devices and printers |
| The computer that serves this analyzer is on a network Windows classes as Public, so Windows Firewall keeps the analyzer out and its results do not arrive. How a network is classed is a security setting for the computer's owner, and the program does not change it: if this is the laboratory's trusted network, make it Private with an administrator's account: Settings > Network & internet > network properties > Private network; or ask the computer's administrator. The program does not open the analyzer's port to public networks: it would then be open on every network the computer joins. | The serving computer is on a Public network. | Ask the computer's administrator to make it Private. | Devices and printers |
| This analyzer's file format is not set yet, so its files will not be read: set it under «Results files» below the card. | The analyzer's file format is not set. | Set it under Results files. | Devices and printers |
| The tube number is digits only: if the analyzer does not read the label and gives the sample its own running number (1, 2, 3...), its result may land on another tube. Turn on barcode reading in the analyzer, or give the tube number a prefix. | The tube number is digits only and the analyzer may number samples itself. | Turn on barcode reading in the analyzer, or give the tube number a prefix. | Devices and printers |
| The file could not be taken: either the library cannot be reached now, or the file does not carry the Clarus signature and was refused. Nothing was changed. | The library could not be reached, or the file lacks the Clarus signature. | Nothing changed; try again later or use a file you hold. | Devices and printers |
| No gateway station is paired with this laboratory. | No station is paired. | Pair one (31.9.6 Task 6 — Pair a new station). | Devices and printers |
| No printer is bound on this station yet — choose one for each role first. | No printer is chosen on the station. | Choose one for each role first. | Devices and printers |
| Every attempt to update the station failed: install the newest version on it by hand. | Every attempt to update the station failed. | Install the newest version on it by hand. | Devices and printers |
41.1.12 Settings, users and setup
| The screen says | Meaning | What to do | Chapter |
|---|---|---|---|
| The picture is larger than 512 KB once it is stored. Choose a smaller one. | The chosen picture is over 512 KB. | Choose a smaller picture. | Laboratory identity |
| Settings saved. | The identity was saved. | Nothing. | Laboratory identity |
| The settings could not be saved. Try again. | The save failed. | Try again. If the server gave a reason, read it (in Laboratory identity it is shown under the save bar); in the Print studio the draft is kept. | Laboratory identity, Messages, Print studio |
| The settings could not be loaded. | The screen could not read the current identity, or the settings did not load. | Retry (use the retry button). | Laboratory identity, Messages |
| The settings could not be loaded. | The server did not answer. | Press the retry button. | General settings |
| The settings could not be saved. Try again. | The save failed. | Try again; if it repeats, check the connection. | General settings |
| Settings saved. | All pending changes were kept. | Nothing. | General settings |
| An example does not give its expected value. Do not sign; contact support. | A calculation check failed. | Do not sign; contact support. | General settings |
| This number's barcode does not fit the label: shorten the format. | The numbering format is too long to print as a barcode. | Shorten the mask or the digits. | General settings |
| The title was switched off, not deleted. | Patients wear the title. | Nothing; it stays for them. | General settings |
| In these branches fewer than two people may release a result… | Turning on separate release would block those branches. | Give a second person the right first. | General settings |
| The language library cannot be reached now. | No internet. | Import a language file. | General settings |
| The current licence allows no more branches. | The licence allows no more branches. | Stop or reuse an existing branch, or ask about the licence. | Branches and linking |
| The branches could not be loaded. | The list did not load. | Press the retry button on the list. | Branches and linking |
| The branch could not be saved. | The branch was not saved. | Read the reason under the form and correct it. | Branches and linking |
| This device does not work until it is licensed: enter its serial number on its «This device» screen. | The device is working but has no licence. | Enter its serial number on its own screen. | Branches and linking |
| This file is not a file from another branch of the laboratory. | The imported file is not from this laboratory's branch. | Import the right file. | Branches and linking |
| This computer is not connected to the laboratory's network. Connect it by cable or Wi-Fi, then check. | The server computer has no network. | Connect it, then press Check now. | Branches and linking |
| Windows classes this network as public, so its firewall keeps the devices out. Make it private on this computer. | Windows classes the network as public. | Press Make this network private. | Branches and linking |
| The certificate names an address this computer no longer has. Renew it, then start again; the devices keep trusting it. | The certificate names an address the computer no longer has. | Press Renew the certificate, then start again. | Branches and linking |
| Clarus Link refused this branch's connection. | Clarus Link refused this branch. | Read the reason under it; it says what to do. | Branches and linking |
| The authority's certificate could not be saved. | The authority's certificate was not saved. | Choose the authority's certificate again, not the key. | Branches and linking |
| The laboratory site could not be loaded. | The data did not arrive. | Press the retry button. | Laboratory site |
| Changes saved. Choose an address for the site to publish it. | Saved, but there is no address yet. | Take a name on Address and publishing. | Laboratory site |
| The site is published at … | The publish succeeded. | Open the address to check. | Laboratory site |
| Could not save or publish. | The save or publish failed. | Read any reason shown beside it and try again. | Laboratory site |
| The picture is larger than 1.5 MB. | The picture is refused. | Choose a smaller one. | Laboratory site |
| The name could not be taken. | The name is unavailable or invalid. | Choose another name. | Laboratory site |
| The domain could not be found. | The check found no such domain. | Check its spelling and its records. | Laboratory site |
| The domain does not point at the website host yet. | The records are missing or not yet visible. | Type the records, wait, and check again. | Laboratory site |
| The preview could not be drawn. | The preview failed. | Press its retry button. | Laboratory site |
| This server does not show the catalogue yet… | The server is older than the screen | Update the program. | Catalogue |
| The catalogue could not be loaded. | A connection fault | Check the connection and reload. | Catalogue |
| The action could not be applied. Nothing was changed. | A bulk change failed | Try again, or check the connection. | Catalogue |
| Only the laboratory's administrator imports. | Your role may not import | Ask the administrator. | Catalogue |
| Only the laboratory's administrator saves reference bands. | Your role may not save bands | Ask the administrator. | Catalogue |
| The server did not accept this band. Check its values. | A limit or age is not valid | Read the red line under the field. | Catalogue |
| The panel was saved, but … members were not accepted | A member does not exist or is itself a panel | Reopen the panel and fix its members. | Catalogue |
| The panel was saved, but its price was not changed. | The price path refused | Read the reason given with it. | Catalogue |
| Could not save. Try again. | The save did not reach the server and gave no reason. | Press the button again; if it repeats, check the connection. | Users and permissions |
| Could not load. | The page's data did not arrive. | Use the retry button shown with it. | Users and permissions |
| A red note at the top of a window (the server's own words) | The server refused: for example a user name already used. | Read it, correct the field, save again. | Users and permissions |
| You do not hold it, so you cannot grant or withdraw it. | The permission is not yours to hand on. | Ask a manager who holds it. | Users and permissions |
| Set for the whole role only… | This permission cannot be given for one branch. | Change it in the roles matrix, as general manager. | Users and permissions |
| Not in force: its giver no longer holds it | The grant's giver lost the permission, so the grant stopped working. | Grant it again from someone who holds it. | Users and permissions |
| The secret's name as kept on the server, not the secret: the program does not take a secret typed here. | No secret is kept yet. | Type the name of the secret kept on the server. | Messages |
| None of them leaves before you choose. | Nothing leaves before your choice. | Choose one of the three buttons. | Messages |
| The design could not be loaded. | The design did not load. | Press the retry button; if it keeps failing, check the connection to the server. | Print studio |
| The preview could not be drawn. | The server could not draw the preview. | Press the retry button, or undo your last change. | Print studio |
| The picture is larger than 700 KB. Make it smaller, then upload it. | The picture is too big. | Make it smaller and upload it again. | Print studio |
| Only a PNG or JPEG picture is taken. | The file is not a PNG or JPEG. | Convert it. | Print studio |
| These elements cover each other on the label: …. Move one of them so that everything reads clearly. | Two elements cover each other on the label. | Drag one away. | Print studio |
| The page has no room for the results if the two spaces together exceed … cm. | The two spaces leave no room for results. | Reduce the spaces. | Print studio |
| The first setup could not be loaded. | The setup could not be loaded. | Press the retry button; check the server is running. | First setup |
| The save failed, so the step stays as it was. | The step could not be saved. | The step stays as it was; try again. | First setup |
| The laboratory's name is not written yet, so the printed report carries no name until it is. | No name is written. | Write Display name in the identity step. | First setup |
| You are alone in the laboratory now. Add an account for everyone who works with you. | Only one working account. | Open the users page and add the others. | First setup |
| The review is recorded. | The review was recorded. | Nothing; go on. | First setup |
| The first setup is finished. | The setup is finished. | Nothing; you return to the home. | First setup |
41.1.13 Data, backups, security and audit
| The screen says | Meaning | What to do | Chapter |
|---|---|---|---|
| Could not load the audit trail. | The list or the box could not be loaded. | Use the retry button on the list; if it persists, check the program is running. | Security and audit |
| No entries match the filters. | No trail entry matches the filters. | Widen the filters or clear the search. | Security and audit |
| No active sessions. | No session is active. | Choose All to see ended ones. | Security and audit |
| No matching events. | No access event matches. | Widen the filters. | Security and audit |
| Export failed. | The export did not complete. | Try again; narrow the filters if the span is large. | Security and audit |
| The access-event chain is broken: contact support | The access events no longer form an unbroken chain. | Do not ignore it; contact support. | Security and audit |
| Choose an entry from the trail to see its details. | No row is chosen. | Click a row. | Security and audit |
| No backup has been taken yet. Take one now. | No backup yet | Press Back up now | Data and continuity |
| Not met: no recent backup | No recent backup | Take one now | Data and continuity |
| The database is back, but the file you chose carries no attachments (pictures and documents). To restore them, choose the backup file whose name ends in .clarus-backup.zip. | The file chosen has no attachments | Choose the .clarus-backup.zip file | Data and continuity |
| We could not read this file. | The file is damaged or half copied | Try another copy | Data and continuity |
| The migration was cut and nothing of it was written. | The program closed during the transfer | Start again | Data and continuity |
| Tick the consent first. | The consent is not ticked | Tick it | Data and continuity |
41.1.14 Licence
| The screen says | Meaning | What to do | Chapter |
|---|---|---|---|
| This is not a valid activation file. | The chosen file is not an activation file. | Choose the file Clarus sent back. | Licence |
| We could not read that file. Please check it and try again. | The file was refused. | Check that it is the right and whole file, and try again. | Licence |
| The request could not be completed. Please try again. | Activation by number failed without a more specific reason. | Check the connection and the card code, and try again; use the file way if there is no internet. | Licence |
| The request file could not be made. | The request file was not saved. | Try again. | Licence |
| The licence could not be checked. | The check did not complete. | Try again later. | Licence |
| The update service could not be reached. | No answer from the update service. | Check the connection and try again. | Licence |
| The notice could not be sent. | The payment notice did not reach the supplier. | Check the connection and try again. | Licence |
| Write the transaction number as the bank app shows it. | The transaction number is missing or too short. | Copy it from the bank application. | Licence |
| The licence could not be moved. | The move failed and nothing was handed back. | Try again; the program keeps working. | Licence |
41.1.15 Printing
| The screen says | Meaning | What to do | Chapter |
|---|---|---|---|
| No printer is bound on this station yet — choose one for each role first. | No printer is chosen on the station. | Choose one for each role in Devices and printers. | Printing |
| Print sits off the label or paper | The paper is not aligned. | In Test page check the printers; press Calibration label to see the label's real edges; adjust the offsets in Printing settings. These corrections never change the PDF a patient receives. | Printing |
| Report print buttons are dim | The order is held. | Read the note under the order; the print buttons come back when the hold is lifted. | Printing |
| The report's PDF could not be downloaded. | The PDF could not be downloaded. | Try again. | Printing |
| The numbering barcode does not fit the label | The numbering format is too long. | Shorten it in the general settings. | Printing |
| Two elements cover each other on the label | The design is crowded. | Drag one away in Print studio. | Printing |
41.2 Questions people ask
Each answer only restates what the chapter named in it says.
41.2.1 Reception
- The order will not save. The screen shows Cannot save: … with the reason, or The order was not saved when the server refused the order: read the reason shown, correct it and save again. See Chapter 5: New order.
- The phone number is refused. A number that is one digit repeated is refused, and so is a number not valid in the laboratory's country. Write the real number, choose No phone, or write a foreign number with + and its country code. See Chapter 5: New order.
- A test is not found, or is found but stopped. No test has that code or name: check the spelling. A test that is off in this laboratory: open the catalogue if you may, or ask whoever manages it. See Chapter 5: New order and Chapter 30: Catalogue.
- The screen says the patient may be registered already. On saving a new file, if a file matches strongly the window A file already matches this patient appears: Use the existing file cancels the save and Open a new file anyway goes on and saves. If only the phone matches, the window This phone number is on another file offers New file: a family member sharing the number and New file: the number moved to someone else. See Chapter 5: New order.
41.2.2 Technologist
- A scanned code finds no sample. The code was mistyped, or it belongs to another branch: scan again, search by the order's number, and check the branch filter. See Samples and Results and approval.
- A value is above the most the analyzer measures. Confirm it, or dilute and rerun. See Results and approval.
- Results of a device are held in Results. A failed run holds that device until the violation is resolved with a written action. See Chapter 12: Daily QC.
- The column of the patient's last result is empty. Your role cannot read patient files; ask the administrator if you need it. See Results and approval.
41.2.3 Approver
- The approval was refused. If the screen says nothing was approved, it is a connection fault: check it and try again. If the server refused some tests, a check failed there; see the red marks in the foot. See Results and approval.
- An order shows a release hold in Delivery. The print buttons stay dim until the hold is lifted: wait, and read the note under the order. A quality-control hold on a device's results is lifted when its violation is resolved with a written action in Daily QC. See Delivery and Daily QC.
- A critical or doubtful value needs a decision. Approving, calling a critical value and withdrawing an approval belong to the role that holds Verify results; amending an approved result also needs Change a result not yet approved. See Results and approval.
41.2.4 Manager
- A branch's money is left out of the overview or the consolidation. No exchange rate is written for its currency in that month: write the rate in Rates of exchange. See Home and Treasury.
- A month cannot be closed. A shift, a recurring expense or a document blocks it: clear the reason listed, then close again. See Chapter 15: Treasury and accounts.
- I cannot decide my own request or purchase order. Another manager decides what you asked, and a purchase order you created or sent is approved by someone else who holds procurement approval. See Chapter 15: Treasury and accounts and Chapter 20: Inventory.
- An order is held back in Delivery because money is left. Collect the remainder, or ask a billing manager. See Delivery.
41.2.5 Administrator
- A new branch cannot be added. The licence allows no more branches: stop or reuse an existing branch, or ask about the licence. See Chapter 27: Branches and linking and Chapter 38: Licence.
- A user cannot grant a permission. A permission can be granted or withdrawn only by someone who holds it. A grant stops working when its giver loses the permission. See Chapter 33: Users and permissions.
- An analyzer sends nothing. Check the cable or network and run the probe; confirm that a station serves it and is marked as cabled to it. See Chapter 31: Devices and printers.
- The screen says there is no recent backup. Take one now. See Chapter 37: Data and continuity.
- An activation file is refused. Choose the file Clarus sent back, and check that it is the right and whole file. See Chapter 38: Licence.
41.3 When a button is dim or missing
A button or a function that is dim or missing almost always means that your role does not hold the permission it needs. The table gives the permission as each chapter names it, and, in the last column, what the chapter says happens to the control without it: it is removed from view, or it stays but dim. This differs from screen to screen. Where a chapter does not say, the cell is empty. A role holds permissions as the administrator set them in Users and permissions; ask the administrator, and do not look for a way around.
| What you want to do | What your role must hold | Chapter | Removed or dim |
|---|---|---|---|
| See the critical button in the top bar (it also needs open critical values) | Verify results | Getting started | removed |
| Use the quick buttons New patient, New order, Receive a sample | Write patient files for the first; Create orders for the other two | Home | |
| See the cards To draw and To receive (not on the approving physician's Home) | Read orders | Home | |
| See the cards To enter and the one for results to approve, and the list of your pending results | Enter results | Home | |
| See the critical values by name | Verify results together with Enter results | Home | |
| See The laboratory at a glance (the figures of the laboratory) | Read analytics | Home | |
| See the panel Waiting by department | Enter results | Home | |
| See Waiting for my decision, Today so far and Clarus services | Manage settings (Today so far also needs Read analytics) | Home | |
| See the money-requests card in Waiting for my decision | Manage billing together with Approve a discount or a refund from the phone | Home | |
| See the card Awaiting a price (on the managers' and reception Homes) | Read orders together with Edit registered orders | Home | |
| See the line to finish the first setup | Manage the laboratory's structure: branches, numbering, first run, external laboratories | Home | |
| Open the order screen and save an order | Create orders | New order | removed (the menu hides the screen) |
| Search for existing patient files and see the matching files (duplicate suggestions) | Read patient files | New order | removed (the duplicate suggestions) |
| Change the price of a test inside the basket | Manage prices, contracts and commissions | New order | removed (the price pencil) |
| Change an external laboratory's cost, make one the test's default, add or register a laboratory | Manage the laboratory's structure: branches, numbering, first run, external laboratories | New order | |
| See the patient's documents strip (existing file only) | See and add the patient's documents | New order | removed |
| Attach the doctor's request | Enter results | New order | dim (the attachment buttons, with a tooltip saying why) |
| Add a doctor from the order (+ add) | Name a referring doctor while ordering | New order | removed (the link) |
| Take payment: the payment controls and the Payment button | Take payments at the desk or Manage billing | New order | dim (with a tooltip saying why; without the right the order is saved with no payment) |
| Give a manual discount | Take payments at the desk or Manage billing | New order | removed (the button) |
| Open a saved order for editing, right after saving it | Edit registered orders | New order | removed (the Edit the order button; the cancel button stays in its place) |
| Open the results of the saved order (Results) | Enter results or Verify results | New order | dim (with a tooltip saying why) |
| Open the barcode board (change which tube holds a test) | Create orders | Samples | removed (its button) |
| See the screen and the list of samples | Read orders or Accept, reject and receive samples | Samples | |
| Open the screen | Enter results or Verify results | Results and approval | removed (the menu hides the screen) |
| Type and save values | Enter results | Results and approval | dim (with a tooltip saying why) |
| Change a value already saved but not yet approved | Enter results and Change a result not yet approved | Results and approval | dim (with a tooltip saying why) |
| Approve; call a critical value; approve the plain ones together | Verify results | Results and approval | dim (with a tooltip saying why) |
| Withdraw the approval of a result | Verify results | Results and approval | removed |
| Amend an approved result | Verify results and Change a result not yet approved | Results and approval | removed |
| Set a test's delta limit | Manage settings | Results and approval | removed (the fields) |
| Share a saved view with a role | Manage settings | Results and approval | removed |
| See the list of results approved today | Publish reports or Verify results | Results and approval | |
| See the Previous column (the patient's last result) | Read patient files | Results and approval | the column says so instead of showing a value |
| Preview the report | Print reports, receipts and labels | Results and approval | removed |
| Open the screen and the page Ready to hand over; hand over, print, publish or stop a link | Publish reports | Delivery | removed (a page you may not use is not shown) |
| Open the page Versions and amendments and sign versions | Publish reports | Delivery | removed (the page) |
| See the page Messages sent and use its two buttons | Publish the portal and communications | Delivery | removed (the page) |
| Collect the remainder from this screen | Take payments at the desk or Manage billing | Delivery | |
| Hand over when the laboratory's setting requires full payment | Manage billing | Delivery | dim (the tooltip reads «The laboratory's setting lets only a billing manager hand over a report with money left») |
| See the directory and read the file's overview | Read patient files | Patients | |
| Create, edit, archive or restore a file; issue a card | Write patient files | Patients | removed (the buttons) |
| Open a new order for the patient; stop or resume uploading the patient's results to the website; send a test to another branch; record the patient's consent to marketing messages | Create orders | Patients | removed (the buttons) |
| Tick the box that marks a file sensitive | The administrator role | Patients | |
| Merge a duplicate into a file | Manage settings | Patients | removed (the button) |
| See the Visits and Attachments tabs | Read orders | Patients | |
| See the Results tab | Enter results or Verify results or Publish reports or Print reports, receipts and labels | Patients | |
| See the Reports tab | Publish reports or Print reports, receipts and labels | Patients | |
| See the Money tab, the Open balance figure and the Balance column of the directory | Take payments at the desk or Manage billing | Patients | |
| See the Samples tab | Read orders or Accept, reject and receive samples | Patients | |
| See the Activity tab | Manage settings | Patients | |
| See and add documents in the Documents tab | See and add the patient's documents | Patients | |
| Send a message (and tick files in the directory for one) | Send messages to patients | Patients | removed (the buttons) |
| Print a report from the file (Print, Print all the patient's reports) | Print reports, receipts and labels | Patients | removed (the buttons) |
| Use Send in the Reports tab, which opens Delivery | Publish reports | Patients | removed (the button) |
| Open the list and see the chosen order | Read orders | Orders | |
| Press New order | Create orders | Orders | removed (the button) |
| Press Edit the order | Edit registered orders | Orders | removed (the button) |
| Press Hand over | Publish reports or Print reports, receipts and labels | Orders | removed (the button; it opens the delivery screen, which only a role that publishes reports can enter) |
| Press Results | Enter results or Verify results | Orders | removed (the button) |
| Open the screen, the queue and the batches (the Waiting for a laboratory window too) | Read orders | Send-outs | |
| Record that a laboratory received a batch | Create orders | Send-outs | removed |
| Enter the returned results; attach the external laboratory's report | Enter results | Send-outs | removed |
| Print the batch's manifest again | Print reports, receipts and labels | Send-outs | removed |
| Open the accounts page | Manage billing or Read analytics | Send-outs | removed |
| Record or void a payment; export the accounts; open the statement of samples sent | Manage billing | Send-outs | removed |
| Print a laboratory's month statement | Manage billing and Print reports, receipts and labels | Send-outs | removed |
| Open the External laboratories page; add, edit, stop or delete laboratories; assign tests; edit price-list lines and routing rules | Manage the laboratory's structure: branches, numbering, first run, external laboratories | Send-outs | removed (the tab is not there) |
| Add a price-list line to the catalogue as a new test | Manage the laboratory's structure: branches, numbering, first run, external laboratories and Manage settings | Send-outs | removed |
| Open the screen and use everything on it (nothing on it is tied to another permission) | Enter results | Daily QC | removed (the menu hides the screen) |
| See the cultures, start one, add susceptibility results, finalize, use the assistant | Enter results | Microbiology | removed (the menu hides the screen) |
| Open the screen, read the drafts, approve, edit, reject, and remove an approved case report | Verify results | Clinical AI | removed (the buttons) |
| Open the Settings window and the Top up window | Manage settings | Clinical AI | removed (the buttons) |
| Open the Ledger window | Manage settings or Manage billing | Clinical AI | removed (the button) |
| Open the screen | Take payments at the desk or Manage billing | Treasury | |
| See the invoices, print a receipt, ask for a refund or for the credit to be handed back, pay an expense recorded as owed | Take payments at the desk | Treasury | removed (a button you may not use is not drawn at all) |
| Change or void an expense, and manage recurring expenses | Manage billing | Treasury | removed |
| See Approval requests | Manage billing and Approve a discount or a refund from the phone | Treasury | removed (a page you do not hold is not in the row of pages) |
| See Shifts and Cash to the centre | Manage billing | Treasury | removed |
| Write off a patient's debt on an invoice | Write off a patient's debt on an invoice | Treasury | removed |
| See Reports | Read the laboratory's books and financial reports | Treasury | removed |
| See The books | Read the laboratory's books and financial reports or Start the books and set the chart, the year and the payment methods' accounts or Close a month of the books | Treasury | removed |
| Start the books, link payment methods to accounts, add or edit tax types | Start the books and set the chart, the year and the payment methods' accounts | Treasury | removed |
| Close a month | Close a month of the books | Treasury | removed (the section is absent from the page) |
| See Receipt vouchers, Bank and Doctors' commissions | Record receipt vouchers | Treasury | removed |
| See Payroll, and add employees, prepare and pay a payroll | Prepare and pay the payroll, and give advances | Treasury | removed |
| Approve a payroll | Reverse an entry of the books and approve a payroll | Treasury | removed (the button is absent) |
| See Suppliers | Record supplier invoices and payments | Treasury | removed |
| See Consolidated and write exchange rates | Read the chain's consolidated books | Treasury | removed |
| Open the screen and use the pages: tests, lists, payers, contracts, plans, commissions, types, and the payers' printed lists | Manage settings | Prices and contracts | removed (the menu hides the screen) |
| Add, change or remove a branch's own price on the page Branch prices | Manage prices, contracts and commissions | Prices and contracts | removed (the buttons are not offered) |
| Open the Payer list library (the button on the page Payers' printed lists) | Manage prices, contracts and commissions | Prices and contracts | removed (the button is not offered) |
| Open the screen; use the queue, stage a claim, export, record answers and transfers, work the statements | Manage billing | Claims | removed (the menu hides the screen) |
| Open Health information exchange for a claim's order | Publish reports | Claims | removed (absent, not dim) |
| Open Payer setup | Manage settings | Claims | removed (the button is not shown, and the address ?page=setup simply opens the queue) |
| Open the screen and read the pages Today, Month, Employees and schedules, Holidays and leave, and Punches without an employee | Read the attendance of every employee | Attendance | |
| Punch by hand, void a punch or a leave, link a number to an employee, save a schedule, add a holiday or a leave, open the Fingerprint devices and Import a file pages | Manage attendance: numbers, schedules, leaves, files, devices and punches by hand | Attendance | removed (the two pages are not offered and the buttons that change anything are absent) |
| Open the screen, read the figures and export them | Read analytics | Follow-up | removed (the menu hides the screen) |
| Choose the whole laboratory instead of one branch (also: not held to particular branches) | Read analytics | Follow-up | removed (the choice The whole laboratory is not offered) |
| Open the screen and see every page and the reports | View inventory and its reports | Inventory | removed (the menu hides the screen) |
| Create or edit items, import items, change lots (quarantine, recall, release), manage suppliers and their prices, manage consumption recipes | Manage inventory items, lots and recalls | Inventory | removed (a button you may not use is not shown) |
| Record a stock movement | Record stock movements | Inventory | removed |
| Start, enter, send for review, post or cancel a stock count | Run and post stock counts | Inventory | removed |
| Write purchase orders, send them for approval, receive, close or cancel them; order from the reorder recommendations | Manage purchase orders and supplier prices | Inventory | removed |
| Approve or refuse a purchase order above the threshold | Approve purchase orders above the threshold | Inventory | removed (also not offered for an order you created or sent yourself) |
| Read all four pages | Enter results or Verify results | Quality | |
| Open, change and close a corrective action; record a competency assessment; add a round and record its result | Verify results | Quality | removed (these buttons do not appear) |
| Open the screen; read the three pages; export the trail; end another person's session; turn access from outside on or off | Manage settings | Security and audit | removed (the menu hides the screen) |
| Open the six pages; create and edit leads, campaigns, offers and sources; issue and stop discount cards | Marketing: campaigns, offers and referrals | Marketing | removed (the menu hides the screen) |
| See the Bulk sending tab, prepare a send and hand it over for sending | Manage settings together with Marketing: campaigns, offers and referrals | Marketing | removed (the tab is not shown) |
| See the screen and its documents | View inventory and its reports or Manage billing or Manage settings or Manage purchase orders and supplier prices | Clarus Market | |
| Update the Market, install from a file, switch the offers on or off | Manage settings | Clarus Market | |
| Apply rows of a price list, and take them back | Manage settings or Manage purchase orders and supplier prices | Clarus Market | |
| Open the screen, change the identity and save it | Manage settings | Laboratory identity | removed (the menu hides the screen) |
| Open the screen and change the settings | Manage settings | General settings | removed (the menu hides the screen) |
| Open the screen and read the branches; add a branch, edit it, set who reaches it, stop or start it | Manage the laboratory's structure: branches, numbering, first run, external laboratories | Branches and linking | |
| Open the laboratory's network to the devices, and give a certificate authority | Manage settings | Branches and linking | |
| See the Branch devices page and allow a new computer's licence number | Manage settings | Branches and linking | |
| See the invitation box, the exchange line and the link controls of a branch on another installation | Manage branches and linking | Branches and linking | removed (they appear only to someone who holds it) |
| See the Samples sent page | Read orders | Branches and linking | |
| Accept, return or cancel a sent sample | Create orders | Branches and linking | |
| Open the screen, edit all four pages, save, publish, take or change the address, and link a domain | Manage settings | Laboratory site | removed (the menu hides the screen) |
| Open the screen and see its filters and the details of a test | Manage settings | Catalogue | |
| Change a test, its intervals and its external laboratories; switch a test on or off; export, import, write a new test, add from the master library; tick tests and act on many at once | Manage settings | Catalogue | removed (the test window is read-only; the bulk tick boxes, the top buttons and the Stop the test switch are absent) |
| Press New panel or package, and open a panel from the Packages page to edit it | Manage the test catalogue together with Manage settings | Catalogue | removed (the button is absent and a row on the Packages page does not open) |
| Type a price in the list or in the panel window | Manage prices, contracts and commissions | Catalogue | removed (the price field is absent from the panel window; in the list the price is plain text) |
| Open the screen and use all four pages | Manage devices | Devices and printers | removed (the menu hides the screen) |
| See the analyzer's messages in full, with the patient's data | Read patient files | Devices and printers | without it the messages are shown with the data hidden |
| Open the screen | Enter results or Verify results or Manage settings | Accreditation and quality | |
| Choose which standards the laboratory follows (Standards); bring standards from the library (From the library) | Manage settings | Accreditation and quality | removed (the buttons are absent) |
| Switch a function of the program on or off | Manage settings | Accreditation and quality | dim (with a line beside it: «Changed by whoever holds the settings permission») |
| Switch a register or the QC hold on or off | Switch the quality functions on or off together with Manage settings | Accreditation and quality | dim (with a line beside it: «Only the laboratory's manager switches it on or off») |
| Judge an item; open, edit, close or void a register entry | Verify results | Accreditation and quality | removed (the judge button and the register buttons are absent; the register buttons are also absent while the register is switched off, whatever your role) |
| Open the screen | Manage user accounts or Manage settings | Users and permissions | |
| Press New user | Manage settings | Users and permissions | removed (a person with only Manage user accounts sees Users without this button) |
| Open the screen, set channels, open or close patients' messages, edit templates and ready messages, decide on waiting messages | Manage settings | Messages | |
| Open the four tabs, change the designs and save them | Manage settings | Print studio | removed (the menu hides the screen) |
| Open the screen, see all five pages and use them (no button on it is tied to a separate permission) | Import data from a previous system or Manage settings | Data and continuity | removed (the menu hides the screen) |
| Open the screen; see the licence, check it, activate it, report a payment, look for an update and move the licence | Manage the licence | Licence | removed (the menu hides the screen) |
| Open the wizard and write its steps | Manage the laboratory's structure: branches, numbering, first run, external laboratories | First setup | removed (the menu hides the screen) |
| Have the wizard notice that the history already came from RealLab | Import data from a previous system | First setup | without it the wizard cannot look, so it always finds none |
| Preview a report on the results screen | Print reports, receipts and labels | Printing | removed (the preview button is not shown) |
| Design the report, receipt and label | Manage settings | Printing |
41.4 See also